Below is a detailed step‑by‑step troubleshooting guide (for a service technician) for the Fresenius 4008 hemodialysis machine. This is a general workflow based on published service/technical manuals and troubleshooting guides. Always refer to the exact service manual version for your machine (software, hardware revisions) before making changes.
Also note: many of these steps require measuring instruments (pressure gauges, conductivity meter, multimeter, dialysis tester, etc.).
⚠ Safety & Preliminary Steps
Disconnect power / ensure safe conditions
Before opening panels or working on internal components, switch off and unplug or isolate power, per safety instructions in the manual.
Observe ESD (electrostatic discharge) precautions when touching PCBs.
Check for active alarms / error messages
Note the error code(s) displayed on the machine.
Clear any user-acknowledged alarms (if possible) so you get a “clean slate.”
Make sure no blocking alarms (e.g. power failure, water alarm) exist before proceeding with diagnostics.
Run a built-in diagnostic / “T1 test”
The 4008 has a built-in self-test procedure (often called T1 test or diagnostics) that checks core subsystems.
If multiple tests fail, enable “Test Service ON” (by adjusting the DIP switch (LP632, e.g. SW4) or service mode) so you can test subsystems individually.
Document which subsystems fail (water path, hydraulics, conductivity, temperature, etc.).
Follow the standard order of troubleshooting
Many guides suggest working errors in this sequence:
Water alarms
Flow alarms
Upper flow alarms
Temperature
Conductivity
Blood leak
Pressure/hydrostatic tests
TMP / continuous ultrafiltration alarms
Hydraulic test fails
Blood pump / arterial / venous / heparin / Diasafe errors
Ultrafiltration (UF) errors
Battery / power supply faults
🔧 Step‑by‑Step Troubleshooting Tasks
Below is a breakdown by subsystem / alarm category, with typical causes and diagnostic steps.
1. Water Alarms / No Water
Symptoms: “No Water” alarm, inability to draw water, low inlet flow.
Possible causes:
Water inlet supply pressure too low (< ~0.95–1.05 bar)
Inlet filter clogged
Inlet valve (V41) malfunction
Pressure regulator fault
Steps:
a. Check municipal water supply: is there adequate pressure and flow?
b. Inspect/replace inlet filter or mesh screen.
c. Test the inlet valve (V41) for proper operation (e.g. measure coil resistance, actuation).
d. Use the diagnostics mode / water path test to trace flow.
If still fails, check pressure regulator or piping blockages.
2. Flow Alarms / Upper Flow Alarms
Symptoms: Flow alarm, inability to reach set flow, or flow too high.
Possible causes:
Blocked filters, tubing, degassing filter
Valve or pump defect
Calibration drift (valves, flow sensors)
Loading pressure incorrect
Steps:
a. Run diagnostics → initiate a flow test (e.g. 300, 500, 800 mL/min) to see how far it deviates.
b. Inspect tubing, lines, degassing filter (clogs, air bubbles).
c. Check valve resistances, actuations.
d. Calibrate flow valves if needed (adjust via calibration menu).
e. For “upper flow alarm,” check if flow is exceeding expected thresholds, possibly due to stuck valve or bypass.
Scribd
3. Temperature Errors / Heater Faults
Symptoms: Temperature alarm, disproportionate temperature readings, heater not functioning.
Possible causes:
Faulty temperature sensor
Heater element failure
Calibration error
Incomplete or failed hot disinfection cycles
Steps:
a. Check incoming water temperature (external reference).
b. Run “Temperature Calibration” routine in service menu; compare reading vs external instrument.
c. Verify heater coil resistance, continuity.
d. Inspect wiring to heater, sensor connectors.
e. If during cleaning/disinfection, verify that hot disinfection program completed properly.
4. Conductivity / Dialysate Mixing Errors
Symptoms: Low or high conductivity alarms, mixing ratio errors.
Possible causes:
Incorrect concentrate or bicarbonate stroke rate
Faulty pumps (concentrate, BIC)
Faulty valves or mixing manifold
Calibration offset / drift
Steps:
a. In diagnostics, check concentrate and bicarbonate pump strokes and suction lines.
b. Check concentrate / BIC pump motors, tubing, check valves.
c. Calibrate mixing ratio (dilution factor).
d. Inspect sensors and wiring for conductivity measurement.
e. Replace faulty concentrate or mixing parts if necessary.
5. Blood Leak Detection Errors
Symptoms: Blood leak alarm (optical or conductivity-based).
Possible causes:
Faulty cuvette / sensor
Contaminated measurement path (residual blood, bubbles)
Calibration of blood leak voltage wrong
Ambient light intrusion
Steps:
a. Inspect the cuvette (glass chamber) for contamination or scratches.
b. Clean or replace cuvette.
c. Calibrate blood leak voltage in service menu (around 5.0 V ± tolerance)
d. Ensure ambient light does not reach the sensor path during test.
e. Reseat connectors and check wiring for the sensor module.
6. Pressure Holding Tests / Hydraulic Failures
Symptoms: Negative pressure alarms, failure of pressure-hold test, TMP/ΔP alarms.
Possible causes:
Leakage in hydraulic circuit
Valve not sealing (V39, V24, V26, etc.)
Faulty pumps (UF pump, flow pump)
Transducer malfunction
Calibration offsets
Steps:
a. Run “Pressure Holding Test” in service/diagnostics menu.
b. If test fails, isolate segments of the hydraulic path (close certain valves) to locate leak.
c. Check valves like V39 (especially SW5 / deep switch relevance).
d. Inspect seals, O-rings, tubing for leakage.
e. Replace defective valves or transducers.
f. Recalibrate pressure sensors (dialysate pressure, etc.).
7. Ultrafiltration (UF) Errors
Symptoms: Insufficient ultrafiltration, UF pump failure, discrepancy between set vs achieved UF.
Possible causes:
UF pump failure or calibration drift
Blocked UF lines or tubing
Spring set or stroke volume incorrect
Error in compensating sensors
Steps:
a. In service mode, check UF pump behavior (stroke count vs expected volume).
b. Inspect UF tubing and check for blockage or kinks.
c. Recalibrate UF pump stroke volume.
d. If spring set or internal mechanical component is worn, replace.
e. Check associated sensors (pressure, flow) that might affect UF control.
8. Blood Pump / Arterial / Venous / Heparin Errors
Symptoms: Blood pump stops, arterial or venous pressure alarms, heparin pump faults.
Possible causes:
Motor issues, worn brushes
Flow sensor or motor coupling mechanical problem
Pressure sensor drift or failure
Tube occlusion or kink
Heparin pump motor or control board failure
Steps:
a. In diagnostics, drive the blood pump in test mode and check motor response.
b. Inspect motor brushes, coupling, wiring.
c. Check pressure sensors on arterial/venous lines; calibrate if necessary.
d. Inspect tubing for occlusions or kinks upstream of pump.
e. For heparin, test heparin pump in diagnostics, verify dosing strokes, inspect motor.
9. Battery / Power Supply Errors
Symptoms: Battery failure / low battery, inability to hold power, power alarms.
Possible causes:
Battery aged or defective
Blown fuse in battery circuit
Fault in 24 V / internal power supply board
Steps:
a. Measure battery voltage under load.
b. Check battery fuse (e.g. 3.15 A fuse)
c. Inspect power supply board (LP647 or equivalent) for faults.
d. Replace battery if it fails capacity test.
📝 Final Steps After Repair / Correction
Re-run the T1 / diagnostic self-test to verify all subsystems now pass.
Clear historic error memory (reset failure record) via the service menu.
Perform calibration routines (flow, temperature, conductivity, pressure) to bring machine back into specification.
Run a complete disinfection / cleaning cycle to confirm no leaks or fluid path errors.
Run validation tests with test fluid (instead of blood) to ensure machine behaves correctly under load.
Log all work done, replaced parts, calibration values for regulatory / maintenance traceability